				Standard type | 
			
				  
				Bottom with parts carrying case | 
		
	
	Apps
	Leakage and pressure test
	Detecting leakage of frictional pressure in pipes
	Working with the flow meter to monitor gas pressure
	Measure wellhead gas pressure and liquid pressure in riser
	Measure steam pressure in energy efficiency tests
	
	Features
	Range: 3...60MPa (type 80) / 0.7...12MPa (80L type)
	Accuracy: 0.03% + 5kPa (type 80) / 0.04% + 1.5kPa (80L type)
	Ergonomics
	All equipment provides a certificate that traces national standards
	Select UKAS certificate for our Pressure Standards Laboratory
	
	Description
	Precision pressure measurement
	Hydraulic pipe tester is easy to carry, can accurately measure any and unstable pressure value, suitable for power plants, pipelines and many other fields.
	Basic principles
	Load the weight onto the piston for the measured pressure value. The downward pressure generated by the weight through the piston rod is balanced with the pressure of the chamber being measured.
	The low pressure precision dial indicates the measured pressure, and the mass of the weight corresponds to the indication on the dial.
	Dragable to national standards
	The weight and piston provided by the pipe tester are calibrated by a standard that can be traced back to the national standard.